4
4- 16 Perfusor® compact, 2.1 gb
2.1
3. Unscrew four screws.
4. Remove cover for drive head housing with toggle, lever and
release shaft (square).
5. Sketch the cable layout.
Note
Pay attention to spring when removing the housing cover.
6. Remove cam switch with pressure spring and pressure pin.
7. Remove push-button sensor board and disconnect plug
connector.
8. Push drive head housing toward the unit.
9. Pull jaws off the drive tube.
10. Pull drive head housing together with clamp off the drive
tube.
11. Assembly is performed in reverse order.
Note
Pay attention to cable layout. The cable must be laid between the
pins and under the light barrier.
Fig.: 4 - 20
Drive head housing Cap for drive head housing
Release shaft
TogglePressure spring
LeverCam switch
Sensor board
Fig.: 4 - 21
Jaws
Drive tube
Axial positioner
Clamp
Drive head housing
Seal washer
Fig.: 4 - 22
Connection cable of push-button senso