Unit Elements
4
Perfusor® compact, 3.0 gb 4- 17
3.0
4.17 Housing Bottom Part, Complete
Designation Ord. No.
Housing bottom part, complete . . . . . . . . . . . . . . . . . . 3450 6594
Exchange
1. Open housing (see "Open unit" ➨ p. 4 - 3).
2. Shift type plate.
a) Warm up type plate with a hair dryer until the adhesive
can be removed (not too hot as otherwise the housing is
damaged).
b) Clean adhesive position on new housing and stick type
plate. New type plates can only be ordered as spare parts
if the old type plates are returned to B.Braun (see
"Ordering of Spare Parts and Test Equipment"
➨ p. 0 - 9).
3. Remove drive from old housing (see "Drive" ➨ p. 4 - 12).
4. Install drive in new housing (see "Drive" ➨ p. 4 - 12).
5. Close housing.
Note
Do not squeeze the cable (see "Close Unit" ➨ p. 4 - 4).
6. Calibrate in Service Program (see "Unit Calibration"
➨ p. 3 - 16).