Checks after Repair
5
5- 2 Perfusor® compact, 2.1 gb
Visual Inspection
1. Check unit for cleanliness, completeness, damage and faults
affecting safety. Pay special attention to the following parts:
- Syringe holder, axial positioner, drive head
- Syringe table and quick reference guide
- Membrane keyboard
- Battery compartment cover, battery compartment and -
contacts
- Unit feet
- MFC connector
- Holder for pole fixation, side snap-in mechanism
- Mains power lead
Functional Inspection
Mechanical Inspection
1. Check function of the holder for pole fixation.
2. Check stacking function of the unit with respect to other
units.
3. Check function of the syringe holder with syringe.
4. Check function of the drive head lock.
Switch on unit.
1. Switch on unit and keep ON-button pressed for max. 20 sec.
Check the screen display during this time. A device alarm is
triggered if the button is actuated for more than 20 sec.
2. The following information appears on-screen when the
button is released:
111.
1
222.
2
555.
5
AA Reference to the instructions
for use (hard- and soft-
ware group)
Last syringe type
3. An audible alarm sounds three times.
2.1