Checks after Repair
5
Perfusor® compact, 2.1 gb 5- 3
Operation
1. Open lock (drive head).
Check push-button sensor alarm. The piston rod symbol must
flash on the LC display if a syringe was not inserted.
2. Insert calibration gauge in the OPS 50 ml slot and close
syringe holder.
The lock must engage automatically and the syringe symbol
in the LC display must not flash.
3. Confirm calibration gauge as OPS with F-button. Program as
OPS syringe type beforehand if necessary.
4. Start pump with a delivery rate of 12.3 ml/h (key sequence 1
2 3).
The pump delivers. The delivery rate set must be displayed.
5. Open lock.
Alarm by buzzer and positive locking sensor alarm. Drive
stops. Lock must not engage in upper position.
6. Plug in MFC service connector and actuate the START button.
Drive delivers at 12.3 ml/h.
7. Change delivery rate to 96 ml/h (key sequence C 9 6 F) during
infusion.
The pump delivers. The delivery rate set must be displayed.
Note
The following staff call modes can be selected in the Service
Program if the unit is switched off: static, dynamic with and
without alarm.
8. Pull syringe holder.
Staff call: red LED in MFC service connector lights up. Drive
stops.
Fig.: 5 - 1 Calibration gauge
Fig.: 5 - 2
OPS 50 ml slot