Brooks Automation
90 278580 Revision A
7 Operation LF80C ZeniD RFID Reader
7.3 Operation of SECS Protocol Product Manual
Description: Result information on the status of the request concerning the
service request.
"NO" Normal operation
Indicates the success of the requested action.
"EE" Execute error
Cannot read tag data. Cannot read ID sequence.
However, equipment is normal.
"CE" Communication error
Syntax error of message or message format or value.
"HE" Hardware error
ID reader/writer head fault, ID reader/writer head is powered off.
"TE" Tag error
Where used S18F2, S18F4, S18F6, S18F8, S18F10, S18F12, S18F14
Description: Indicates an action to be performed by the subsystem.
Used to differentiate between the different subsystem commands indicated.
"ChangeStatus" Change status
"GetStatus" Get status
"Reset" … Reset CIDRW
"PerformDiagnostics"… Perform diagnostics
"DefaultParams" Reset default parameter*
"FactorySettingParams" … Reset factory settings and default parameter
ADJUST“ Automatic antenna calibration (parameter 33)
"SetLED“ … Set LED*
Where used S18F13
SSACK Format: A[2]
SSCMD
Format: A[max
18]