Measuring Ampifier Type 2525
User Manual Vol.2
5–7
BE1394– 11
Chapter 5 – General Control Messages
Status and Configuration Check
The query Measurement_Status_Enable? returns the value or options set, as shown
in Fig.5.5.
Bit Function Value
None 0
1 Common_Mode_Overload 1
2 Signal_Overload 2
3 ALArm 4
4 Deltatron_Error 8
5 Upper_20_Db16
6 Auto_Ranging 32
7 Detector_Overload 64
8 not defined 128
ALL 255
Table 5.2 The values contained in the Measurement_Status event
enable register
Fig.5.5 Syntax for Measurement_Status_Enable? query and amplifier response
941131e
Te
CHANNEL_1
CHANNEL_1
MEASUREMENT_STATUS_ENABLE?
To 2525
:
Header OFf
Te
From 2525
: :
Value
NR1
SP
—
:
MEASUREMENT_STATUS_ENABLE