E. Customer Acceptance Test Protocol
DOC-M88-EXX157_V17_04.2019 61 / 96
Instrument Configuration D8 DISCOVER
Instrument Details Set Values for Validation
Diffractometer D8 DISCOVER Tube Power: 40kV/40mA (for FL tubes)
Tube Power: 40kV/ 30mA (for FF tubes)
TXS Power: 50kV/100mA (for 0.3x3mm
2
filament, line and spot focus)
TXS Power: 50kV/22mA (for 0.1x1mm
2
filament, spot focus)
Vertical
goniometer
Theta/Theta
Theta/2Theta
Height:
150
214
258
Radius:
Primary radius
Secondary radius
....................[mm]
....................[mm]
Horizontal
goniometer
Theta/2Theta
Height:
214
258
Radius:
Primary radius
Secondary radius
....................[mm]
....................[mm]
Sample stage Standard
Rotating
FLIP-STICK Sample position 5
AUTO CHANGER Sample position A1
Compact Cradle/ Compact
Cradle
Plus
Centric Eulerian Cradle
UMC 150/151/1516/150 HTC
Compact UMC
Monochroma-
tisation
Kβ-Filter
Secondary monochromator
Focusing monochromator
Discriminator Settings
(LYNXEYEXE/ LYNXEYEXE‑T)
Copy values from High Resolution mode:
Low Thr. [V] High Thr. [V] Veto Thr. [V]

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bruker D8 Series and is the answer not in the manual?

Bruker D8 Series Specifications

General IconGeneral
ManufacturerBruker
SeriesD8
X-ray SourceSealed tube or rotating anode
GoniometerVarious types available, including θ/θ, θ/2θ, and vertical goniometers
Sample StageVarious options including XYZ stages and sample changers
SoftwareDIFFRAC.SUITE

Related product manuals