E. Customer Acceptance Test Protocol
62 / 96 DOC-M88-EXX157_V17_04.2019
Instrument Details Set Values for Validation
Slits Fixed divergence slit Divergence slit: 0.3° (approx. 0.6mm)
Fixed antiscatter slit Antiscatter slit: 0.3° (approx. 0.6mm)
Variable divergence slit Divergence slit: 0.3° (approx. 0.6mm)
Variable antiscatter slit Antiscatter slit: 0.3° (approx. 0.6mm)
Motorized air scatter screen Automatic mode
Primary axial Soller slit Primary axial Soller slit: 2.5°
Secondary axial Soller slit Secondary axial Soller slit: 2.5°
Receiving slit Receiving slit 0.1mm
Detector
(See User
Manuals)
Scintillation counter
LYNXEYE, SSD160
LYNXEYEXE
LYNXEYEXE‑T
LYNXEYE‑2
SSD160‑2
LYNXEYE User Manual DOC-M88-EXX095
LYNXEYEXE User Manual DOC-M88-EXX240
LYNXEYEXE‑T User ManualDOC-M88-EXX239
LYNXEYE‑2 User Manual DOC-M88-EXX305
SSD160-2 User Manual DOC-M88-EXX306
Solid state detector
(SOL‑XE)
SOL‑XE User Manual DOC-M88-EXX113
VÅNTEC‑1 VÅNTEC‑1 User Manual DOC-M88-EXX072
EIGER2R500K EIGER2R500K User Manual DOC-M88-EXX293
Comments