Technical Data
E1400912_1_001 19
4 Technical Data
4.1 General Information
No. Series Type Weight
(kg)
Length
(cm)
Width
(cm)
Heigth
(cm)
1 mq20 Toothpaste
Analyzer
105 65 75 33
Table4.1: Technical Data: General Information
All specifications above are without a PC. Weights and dimensions apply to the instruments,
not to the packaged systems.
4.2 Connection Values
Grounding is vital for the safe operation of the instrument. Never operate with the grounding
conductor disconnected. If the local wiring does not provide a ground, an earth-to-chassis
connection must be added at the user’s location.
Note: Since the input voltage may be unstable in a laboratory environment, it is
recommended to connect the minispec to a separate circuit that is not used by any other
equipment. Furthermore, a voltage stabilizer or UPS may be required to avoid instrument
instabilities or other unexpected system behavior caused by a fluctuating input voltage.
No. Series Type ND/NE/NF Voltage/VAC Apparent
Power
Consumption,
Maximum/VA
Circuit
Protection
(Two
indications:
NF/ND Series)
Freque
ncy/Hz
1 mq20 Toothpaste
Analyzer
NF 100-240 300/250 3.15/2.5 AT (2
pieces)
50-60
Table4.2: Electrical Connection Values