EasyManuals Logo
Home>Bruker>Measuring Instruments>minispec mq20

Bruker minispec mq20 User Manual

Bruker minispec mq20
86 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #35 background imageLoading...
Page #35 background image
Running the minispec mq20 Toothpaste Analyzer
E1400912_1_001 35
During Daily Check, the system runs a number of tests, all following one after the other.
If all the tests passed, the software will indicate this accordingly. The corresponding message
will also be displayed in the main menu.
7.3 Calibrations
TD-NMR is a secondary technique and as such the Toothpaste Analyzer needs to be
calibrated with reference standards.
For all other applications the customer will need to calibrate the system with their own
reference standards. Typically 4-6 standards with known reference values are needed per
calibration line. In most cases the mq20 Toothpaste Analyzers are calibrated in a classical
way, i.e. the NMR signals of FID or Hahn-echo (or both) are considered and the
corresponding amplitudes are correlated with the sample properties.
For different types of samples (like different types toothpaste or mouthwash) the user may
need to create different calibration lines, one line per type of sample. It is recommended to
name the calibrations accordingly.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bruker minispec mq20 and is the answer not in the manual?

Bruker minispec mq20 Specifications

General IconGeneral
BrandBruker
Modelminispec mq20
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals