SYSTEM SETTINGS
Average RR/QTcB/QTcF
Enabling this option will allow the following to appear on the report:
an averaged RR value.
a Bazett’s corrected QT value along with the default linear QTc value.
a Fridericia corrected QT value along with the default linear QTc value.
ECG Capture
Defines whether or not the ELI 280 will automatically display the Best 10 seconds of data acquired or the last 10
seconds of data acquired.
NOTE: The ECG Capture Mode can also be changed for a single ECG by touching anywhere on the ECG
waveform in the acquired view.
Pace Spike Channel
This control allows the user to decide and set as a default, whether a pacemaker spike notification marker will
display at the base of the ECG printout. A pacemaker spike notification marker coincides with each pacemaker
event.
NOTE: The Pacemaker Spike Channel setting can also be toggled on or off for a single ECG by touching
anywhere on the ECG waveform in the acquired view.
ECG Display Speed
This control allows the user to set the default display speed to 5 mm/s, 10 mm/s, 25 mm/s or 50 mm/s for ECG
viewing.
NOTE: The Display Speed can also be changed for a single ECG by touching anywhere on the ECG
waveform in the real time view.
ECG Print Speed
This control allows the user to set the default paper speed to 25 mm/s or 50 mm/s for ECG printouts.
NOTE: The Print Speed can also be changed for a single ECG by touching anywhere on the ECG
waveform in the acquired view.
Number of Copies
This utility allows the user to select the number of printed copies that automatically print when an ECG is taken. A
zero (0) setting prints no copies. Selecting one (1) prints the original copy, two (2) prints the original plus one copy,
and so on up to 9 copies.
Copies with Interpretation
This utility allows the user to select the number of printed copies containing interpretation when an ECG is taken. A
zero (0) setting prints the first ECG with interpretation and all consecutive copies up to nine (9) printed without
interpretation. Settings from one (1) to nine (9) contain an ECG interpretation on that number of selected printed
copies. All copies display patient demographics and measurements.