35
Adjustment and commissioning
Type 8202/8222
9.7. Choosing the output wiring
mode
The setting has no effect on a version with one fixed
connector, if the sole current output is wired.
Param
This is
when the
device is be-
ing parame-
tered............
....................
This is
when the
device is be-
ing parame-
tered............
....................
Outputs
HWMode
sink/NPN
source/PNP
9.8. Setting the probe parameters of
the 8202
The monitoring of the redox ("ORP") sensor is impossible if
the measurement mode is set to "asymmetrical".
This is
when the
device is be-
ing parame-
tered............
....................
Param
Sensor Type: pH
ORP
This is
when the
device is be-
ing parame-
tered............
....................
Mode:
Symmetric
Asymmetric
Mains Fcy:
50Hz
60Hz
9.9. Activating/deactivating the Hold
function
The Hold mode is automatically deactivated when the trans-
mitter restarts after a power interruption, if the Hold mode
was activated at the moment of the power cut-off.
Calib System Hold
Hold:Disable
Hold:Enable
English