36
Adjustment and commissioning
Type 8202/8222
9.10. Calibrate the pH or redox probe (8202)
danger
Risk of injury due to electrical voltage.
Observe all applicable accident protection and safety regulations for electrical equipment.
Risk of injury due to the nature of the fluid.
Respect the regulations on accident prevention and safety relating to the use of aggressive fluids.
Choose the type of probe the transmitter 8202 is fitted with before calibrating the sensor.
In order not to interrupt the process, activate the HOLD function. See chap. „9.9“.
Before each calibration, correctly clean the electrodes with a suitable product.
In a 2-point calibration, the buffer solutions used must be at the same temperature.
Set the periodicity of calibrations in the "Interval" function in the sub-menu "Calib interval" (see previous page): each time a calibration
is due, the transmitter generates a "maintenance" event and a "warning" message.