Adjustment and start-up
9.11.5 Setting the data displayed in the Process level
See chpt. 9.9 to access the Parameters menu.
CondS
TDSppm
Filter:
None
Fast
Slow
EnabledLine1 / Line2:
PVar:
Disabled
CondR
TempC
TempF
µS/cm
°C
Unit:
mS/cm
°F
If PVar = "TempC"
If PVar = "TempF"
If PVar = "CondS"
If PVar = "CondS"
W.cm
kW.cm
If PVar = "CondR"
If PVar = "CondR"
M
W.cm
If PVar = "CondR"
ppm
If PVar = "TDSppm"
Display
This is
when the
device is be-
ing parame-
tered............
....................
Param
This is
when the
device is be-
ing parame-
tered............
....................
PVAR: choose the process value to be displayed in the line selected.
UNIT: choose the unit for the process value displayed.
FILTER: choose the filter level for the measurement values displayed on the line selected. Three filter levels are
proposed: “slow”, “fast” or “none”.
"slow" "fast" "none"
150 ms
6 s
t
30 s
t
t
Fig. 31 : Filter curves
English
Type 8222 ELEMENT