EasyManua.ls Logo

Burkert ELEMENT 8222 - Page 54

Burkert ELEMENT 8222
72 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
54
Adjustment and start-up
Calibrate the conductivity sensor (“Calibration” function in the “Probe” menu)
Calibration consists in determining the C constant specific to each conductivity sensor using a solution with a
known conductivity.
In order not to interrupt the process, activate the HOLD function. Refer to chpt. 9.12.1.
Before each calibration, correctly clean the electrodes with a suitable product.
Set the periodicity of calibrations in the "Interval" function in the sub-menu "Calib interval": each time a
calibration is due, the device generates a "maintenance" event and a "warning" message.
OK
Save or not the calibration result by
choosing "Yes" or "No".
The message "Error: out of range" signals that
the cell constant is out of the authorized range
(< 0.008 or > 12); this may be due to either:
a mistake made when entering the conductivity,
or
the conductivity sensor, which is not able to
measure the solution conductivity.
BACK
Calib Sensor CalibrationProbe
Immerse the clean conductivity sensor in the solution with a known
conductivity; the device alternately displays:
the measured temperature of the solution
the measured conductivity of the solution
Enter the conductivity, at
the fluid temperature, of
the reference solution used
(marked on the bottle or
measured using a reference
instrument).
Change the unit if necessary.
Calibration
5.023 µS/cm
5.000 µS/cm
Cal. Result
+1.00000
The device displays the cell
constant as calibration result.
OK
Cal. Result
Save: Yes/No
"Error: out of
range"
Probe
English
Type 8222 ELEMENT

Table of Contents

Related product manuals