10
General
Description
Section 1: General Description
Lock Level Table
is table lists the operations that are accessible in each lock level while
the pump is stopped and running. LL0 permits complete access to all
programming and operating functions. LL1 permits limited control of
pump programming and operations. LL2 permits only minimal control
of pump operations.
*Continuous Delivery Mode = CDM; Intermittent Delivery Mode = IDM
Pump Operations
and Programming
Stopped Running
LL0 LL1 LL2
Any Lock
Level
Stop/Start the pump Yes Yes Yes Yes
Reset Reservoir Volume Yes Yes Yes No
Prime Yes Yes No No
Change the Lock Level
Yes, w/
code
Yes, w/
code
Yes, w/
code
No
Change Continuous Rate
(CDM)*
Yes
Up to LL0
value
No No
Clear Given amount
(CDM & IDM)*
Yes Yes No No
Change Dose Volume (IDM)* Yes No No No
Change Dose Duration (IDM)* Yes No No No
Change Dose Cycle (IDM)* Yes No No No
Change KVO Rate (IDM)* Yes No No No
Change Dose Starts in Delay
(IDM)*
Yes Yes No No
Biomed Functions
Access to Functions
Yes, w/
code
No No No
Air Detector On/O
Yes, w/
code
View Only View Only View Only
Upstream Occlusion Sensor
On/O
Yes, w/
code
View Only View Only View Only
Changing the Delivery Mode
Yes, w/
code
View Only View Only View Only

Table of Contents