57
Reference
Section 5: Reference
IEC 61000-4-4 (2004), Electromagnetic Compatibility (EMC), Part 4-4:
Testing and measurement techniques. Electrical fast transient/burst im-
munity test.
IEC 61000-4-5 (2001), Electromagnetic Compatibility (EMC), Part 4-5:
Testing and measurement techniques. Surge immunity test.
IEC 61000-4-6 (2001), Electromagnetic Compatibility (EMC), Part 4-6:
Testing and measurement techniques. Immunity to conducted disturbanc-
es, induced by radio-frequency elds.
IEC 61000-4-8 (2001), Electromagnetic Compatibility (EMC), Part 4-8:
Testing and measurement techniques. Power frequency magnetic eld im-
munity test.
IEC 61000-4-11 (2004), Electromagnetic Compatibility (EMC), Part 4-11:
Testing and measurement techniques. Voltage dips, short interruptions and
voltage variations immunity test.
CISPR11 (1997), Limits and methods of measurement of electromagnetic
disturbance characteristics of industrial, scientic and medical (ISM) radio
frequency equipment. Amendment 1 (1999) Amendment 2 (2002).
Specications (Nominal)
General Pump Specications
CADD
™
Medication Cassette Reservoir P/N 21-7002 and CADD
™
Extension
Set P/N 21-7045 were used to test the pump.
Resolution ..............................Medication Cassette Reservoir or CADD™ Admin-
istration Set, 0.050 ml per pump stroke nominal
Size .......................................4.1 cm × 9.5 cm × 11.2 cm (1.6 in. × 3.8 in. ×
4.4 in.) excluding cassette or other accessories
Weight .................................392 g (13.8 oz.) including 2 AA batteries, empty
100 ml Medication Cassette Reservoir, and
excluding other accessories
Classication
(IEC 601-1) ..........................CF , Class II K
Moisture Protection ..........Splashproof (E)