58
Reference
Section 5: Reference
Pump Alarms .....................Low battery power; depleted battery power; bat-
tery dislodged; pump stopped; pump fault; low
reservoir volume; high delivery pressure; air in
line; disposable not attached when run attempt-
ed; motor locked; upstream occlusion; reservoir
volume empty; program incomplete; key stuck;
disposable detached.
Maximum Infusion
Pressure ..............................40.0 psi [2.76 bar]
Maximum Time to
Occlusion Alarm ...............0.1 ml/hr: 2.0 hours
24 ml/hr: 78 seconds
Bolus Volume at
Occlusion Alarm
Pressure ...............................0.1 ml/hr: <0.15 ml
24 ml/hr: 0.44 ml
Power Sources .......................Two AA (IEC LR6) alkaline batteries; AC Adapter.
e expected life of 2 AA batteries is 15 hours
at 125 ml/hour, or approximately 14 days at
10 ml/day (nominal). is estimate is based on
laboratory tests conducted at room temperature
using 2 new batteries. Actual battery life will
vary depending on the brand of battery, battery
shelf life, temperature conditions, and delivery
rate. It is recommended that 2 new AA batteries
be kept available for replacement if necessary.
An internal battery powers the clock. When
it is depleted, it cannot reliably maintain the
clock time. is battery must be replaced by
the manufacturer. e internal battery has an
expected life of 5 years.
System Operating
Temperature .......................+C to 40°C (35°F to 104°F)
System Storage
and Transportation
Temperature .......................-20°C to 60°C (-4°F to 140°F)
When shipping pump, use pump case.