Section 2: Pump Setup and Programming
Pump Setup &
Programming
37
Inserting the Tubing into the Air Detector
WARNING: When the Air Detector is turned o, the pump will not
detect air in the uid path. Periodically inspect the uid path and
remove any air to prevent air embolism. Air embolism could result in
death or serious injury to the patient.
(See Section 4, Biomed Functions, for instructions on how to turn the air
detector on and o.)
1. If the Air Detector is in use, make a small loop of tubing under-
neath the air detector and hold it with your thumb.
2. Place the tubing over the groove in the air detector and tuck it
under the catch.
Catch