Section 4: Biomed Functions
45
4.0 Biomed Functions
Overview: Accessing the Biomed Functions
e Biomed Functions are pump congurations that are less frequently
changed. e Biomed Functions are accessible only when the pump is
stopped and in Lock Level 0.
To Access the Biomed Functions
1. Press
Œ
. e current lock level will appear.
2. Press
Œ
or
¤
. “Code 0” will appear.
3. Press
´
or
Î
until ** Text Omitted **
WARNING: Do not disclose to the patient the pump’s security
codes or any other information that would allow the patient
complete access to all programming and operating functions.
Improper programming could result in death or serious injury
to the patient.
4. Press
to select the setting you wish to view or change, then
follow the instructions in this section for the appropriate screen.
NOTE: To leave a Biomed Function unchanged, press
.
5. To exit the Biomed Functions, press
until you get to the
screen that reads, “NEXT for Biomed, ENTER for main.
6. Press
¤
to return to the main screen.