62
Reference
Section 5: Reference
Accuracy Test Results
e following graphs are designed to show ow accuracy of the infusion
system plotted against given time periods.
Flow rate: Intermediate
Time Interval: 0.5 min
Total Time: 120 min
Programmed Rate: 24.0 ml/hr
Trumpet curve: Intermediate rate
Programmed Rate: 24.0 ml/hr
Average Flow Rate: 23.9227 ml/hr
Mean Flow Error: -0.32%
Flow
(ml/hr)
T (min)
0 12012010080604020 1109070503010
0
12
24
36
48
Flow
(ml/hr)
T (min)
% Error
of Flow
T (min)