EasyManua.ls Logo

CADD Legacy Plus 6500 - Accuracy Test Results

CADD Legacy Plus 6500
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
62
Reference
Section 5: Reference
Accuracy Test Results
e following graphs are designed to show ow accuracy of the infusion
system plotted against given time periods.
Flow rate: Intermediate
Time Interval: 0.5 min
Total Time: 120 min
Programmed Rate: 24.0 ml/hr
Trumpet curve: Intermediate rate
Programmed Rate: 24.0 ml/hr
Average Flow Rate: 23.9227 ml/hr
Mean Flow Error: -0.32%
Flow
(ml/hr)
T (min)
0 12012010080604020 1109070503010
0
12
24
36
48
Flow
(ml/hr)
T (min)
% Error
of Flow
T (min)