Accuracy Tests
Gravimetric Accuracy Testing
Description
A Medication Cassette Reservoir is partially lled with water and
weighed, then attached to a pump that is set to deliver a certain
amount of water. e Medication Cassette Reservoir is then removed
and weighed again. e amount of water delivered is compared to the
amount that the pump should have delivered.
Nominal system accuracy is given in the technical specications section
for the pump. at is, under the test conditions described below, the ac-
curacy of the pump and Medication Cassette Reservoir will be nominal
with a 90% condence level. e nominal test conditions are as follows:
degassed water at 25 ± 5°C without back pressure.
Equipment needed
50 or 100 ml Medication Cassette Reservoir with attached CADD
Extension Set with Anti-Siphon Valve OR
50 or 100 ml Medication Cassette Reservoir with ow stop feature
with attached CADD Extension Set (catalog numbers start with
21-73xx)
50 or 60 ml syringe
A balance accurate to 0.1 g
40 ml of room temperature water
Procedure
1. Fill the 50 or 60 ml syringe with 40 ml of water. Transfer the water
into a Medication Cassette Reservoir.
2. Remove any air from the Medication Cassette Reservoir by aspi-
rating the air with the syringe. Attach the CADD Extension Set.
Prime the tubing so it is lled with uid to the end of the extension
set Luer.