Section 2: Pump Setup and Programming
Pump Setup &
Programming
39
Setting the Lock Level for the Patient
e Lock Level must be changed to LL1 or LL2 to prevent the patient from
having complete access to all programming and operating functions.
NOTE: You may change the lock level at any time by stopping the pump
and following the procedure below.
To change the lock level
1. Press
Œ
.
2. e current lock level will appear.
3. Press
´
or
Î
until the desired lock level (LL1 or LL2) appears.
4. Press
Œ
again or
¤
. “Code 0” will appear.
5. Press
´
or
Î
until ** Text Omitted **
6. Press
Œ
or
¤
to set the new lock level.
WARNING: Do not disclose to the patient the pumps security
codes or any other information that would allow the patient
complete access to all programming and operating functions.
Improper programming could result in death or serious injury
to the patient.