Section 4: Biomed Functions
48
Changing the Delivery Method
Delivery can be changed between Continuous Delivery and Intermit-
tent Delivery. is allows for either continuous medication delivery or
infusion of a prescribed volume of a medication over a specied time
period. When the delivery method is changed, programming will revert
to default settings.
1. With the pump stopped and in LL0, access Biomed Functions.
(Refer to the beginning of the Biomed Functions section for instruc-
tions on how to access Biomed Functions.)
2. Press
until “Delivery Mode” appears.
3. Press
´
or
Î
to select “Continuous” or “Intermittent.
4. Press
¤
to enter the change.