Section 2: Pump Setup and Programming
Pump Setup &
Programming
40
CDM: Programming with Upper Limits, Adjusting Rate
in Lock Level 1
is feature may be used in the continuous delivery mode only. If a
prescription allows for the Continuous Rate to be adjusted during the
course of therapy, you may wish to operate the pump in LL1. en, when
necessary, you can adjust the Continuous Rate up to the maximum value
that was programmed in LL0.
Programming the pump to use this feature
1. During initial programming in LL0, enter the upper limit value
for the Continuous Rate. (is will be the maximum value when
the pump is in LL1.)
2. Aer you are nished programming, change the lock level to LL1.
3. Decrease the Continuous Rate to its starting value, then press
¤
.
Adjusting the rate while the pump is in use
If it becomes necessary to increase the Continuous Rate during the
course of therapy, stop the pump but remain in LL1.
1. Press
until the Continuous Rate screen appears.
2. Press
´
or
Î
to select the desired value, then press
¤
.
3. Restart the pump if appropriate.