Section 2: Pump Setup and Programming
Pump Setup &
Programming
16
Watching Power Up
When you install the batteries, the pump will start its power up sequence
during which it performs self-tests and displays programmed values.
Watch for the following:
Pump model number and serial number appear unless an error has
occurred, then the last error code (“LEC”) if any, will appear.
e soware version will appear.
e display will turn on, showing a series of blocks. Look for any
blank areas, which would indicate a faulty display.
e display will turn o briey.
e pump’s program screens will appear, followed by screens showing
the Air Detector status, Upstream Occlusion sensor status, and lock
level setting. e pump will beep aer each screen. If messages ap-
pear, see the Messages and Alarms Table in Section 5 of this manual
for further explanation and instruction.
When power up is complete, 6 beeps will sound, and the pump will be
stopped on the main screen.
NOTE: To move quickly through the power-up screens, press
repeatedly. To skip the automatic review entirely, press
Î
. If you attempt
to skip screens before the pump is powered up, it will not respond.