Section 4: Biomed Functions
46
Air Detector On/O
e Air Detector screen can be set to On-High, On-Low, or O.
WARNING: When the Air Detector is turned o, the pump will not
detect air in the uid path. Periodically inspect the uid path and
remove any air to prevent air embolism. Air embolism could result in
serious injury or death to the patient.
1. With the pump stopped and in LL0, access Biomed Functions. (Re-
fer to the beginning of the Biomed Functions section for instruc-
tions on how to access Biomed Functions.)
2. Press
„
until “Air Detector” appears.
3. Use
´
or
Î
to select On-High, On-Low, or O.
• On-high is the highest sensitivity, where the smallest bubbles will
be detected.
• On-low is lower sensitivity, where only the larger bubbles will be
detected. See Specications in Section 5.
4. Press
¤
to enter the change.
5. Press
„
to go to the next screen.