Messages and Alarms, Alphabetical List
Messages and Alarms Description/Corrective Action
Air In Line Detected
Two-Tone alarm
e Air Detector has detected air in the
uid path; the uid path may contain air
bubbles, or the tubing may not be fully
threaded through the Air Detector. Press
⁄
or
„
to silence the alarm, then:
•
Make sure the tubing is threaded
properly.
•
If the uid path contains air bubbles,
close the clamps and disconnect the uid
path from the patient. en follow the
instructions for removing air by priming
the pump, described in Section 2. Restart
the pump.
Battery Depleted
Two-Tone alarm
e battery power is too low to operate the
pump. e pump is now stopped.
•
Change the batteries immediately.
•
Press and hold
⁄
to restart the pump.
Battery Removed
Pump won’t run
Two-Tone alarm
With the AC adapter attached, the AA
batteries have been removed while the pump
is running, or you have tried to start the
pump with depleted batteries. e pump is
now stopped. Press
⁄
or
„
to silence
the alarm. Reinstall batteries or install new
batteries. Press and hold
⁄
to restart the
pump.
Error
Two-Tone alarm
An error has occurred. Remove the pump
from service and contact Customer Service
to return the pump for service.