Messages and Alarms, Alphabetical List
Messages and Alarms Description/Corrective Action
Air In Line Detected
Two-Tone alarm
e Air Detector has detected air in the
uid path; the uid path may contain air
bubbles, or the tubing may not be fully
threaded through the Air Detector. Press
or
to silence the alarm, then:
Make sure the tubing is threaded
properly.
If the uid path contains air bubbles,
close the clamps and disconnect the uid
path from the patient. en follow the
instructions for removing air by priming
the pump, described in Section 2. Restart
the pump.
Battery Depleted
Two-Tone alarm
e battery power is too low to operate the
pump. e pump is now stopped.
Change the batteries immediately.
Press and hold
to restart the pump.
Battery Removed
Pump won’t run
Two-Tone alarm
With the AC adapter attached, the AA
batteries have been removed while the pump
is running, or you have tried to start the
pump with depleted batteries. e pump is
now stopped. Press
or
to silence
the alarm. Reinstall batteries or install new
batteries. Press and hold
to restart the
pump.
Error
Two-Tone alarm
An error has occurred. Remove the pump
from service and contact Customer Service
to return the pump for service.