Section 2: Pump Setup and Programming
Pump Setup &
Programming
18
Programming the Pump: General Instructions
e procedure for changing a programmed setting is similar for most
programming screens.
WARNING: Ensure that the ± 6% System Delivery Accuracy specica-
tion is taken into account when programming the pump and/or lling
the Medication Cassette Reservoir. Failure to do so may result in
medication in the reservoir becoming depleted sooner than expected.
If the pump is being used to deliver critical or life sustaining medica-
tion, the interruption in the delivery of medication could result in
patient injury or death.
Make sure the pump is stopped and in Lock Level 0.
To begin programming, start at the main screen and press
.
To change a setting, press
´
or
Î
until the desired setting ap-
pears. (Press and hold these keys to change values with increasing
speed.)
Press
¤
within 25 seconds to conrm a change or the screen will
revert to the previous setting.
If any key other than
¤
is pressed, “Value not saved” will appear.
Press
to return to the screen being programmed, scroll to the
desired value, and press
¤
.
Press
to advance to the next screen.
To leave a setting unchanged, press
to go to the next screen.