Section 2: Pump Setup and Programming
Pump Setup &
Programming
29
Programming Intermittent Delivery
WARNING: is pump delivers medication at a continuous rate or
intermittent rate in milliliters per hour (ml/hr). Programming the
pump at a delivery rate other than what is prescribed will cause over
or under delivery of medication, which could result in patient injury
or death.
Be sure the pump is in the intermittent delivery mode versus the contin-
uous delivery mode. To program the pump, enter the prescribed values.
1. Begin at the main screen.
• Make sure the pump is in LL0.
• Make sure STOPPED appears on the main screen.
• Press
„
to begin.
2. Enter the Reservoir Volume.
• Press
´
or
Î
to select the volume of a lled uid container.
(If you do not wish to use the Reservoir Volume feature, scroll
down to “Not In Use” located before 1.)
• Press
¤
.
• Press
„
.
3. Enter the Dose Volume.
• Press
´
or
Î
to select the desired volume.
• Press
¤
.
• Press
„
.
NOTE: e maximum rate of the pump is 125 ml/hr. e pump will
only allow a dose duration that will not cause a rate greater than
125 ml/hr for the programmed volume.
4. Enter the Dose Duration.
• Press
´
or
Î
to select the desired duration.
• Press
¤
.
• Press
„
.