Section 2: Pump Setup and Programming
Pump Setup &
Programming
30
5. Enter the Dose Cycle.
• Press
´
or
Î
to select the desired cycle.
• Press
¤
.
• Press
„
.
6. Enter the KVO rate.
• Press
´
or
Î
to select the desired KVO rate.
• Press
¤
.
• Press
„
.
7. Acknowledge Dose Remaining (if applicable).
• Press
„
.
8. Enter a delay or “Immediate” for Dose Starts in _____ .
• Press
´
or
Î
to select the desired amount of delay until the
next dose (or “Immediate”).
• Press
¤
.
• Press
„
.
9. Clear Given.
• Press
¤
if you wish to clear the amount given.
• Press
„
.
10. Verify the Air Detector status.
• Make sure the desired setting is displayed. is screen will show
whether the Air Detector is turned on (high or low) or o.
WARNING: When the Air Detector is turned o, the pump will
not detect air in the uid path. Periodically inspect the uid
path and remove any air to prevent air embolism. Air embolism
could result in death or serious injury to the patient.
• If you need to change the Air Detector setting, see Section 4,
Biomed Functions.