Section 2: Pump Setup and Programming
Pump Setup &
Programming
31
Press
.
11. Verify the Upstream Sensor status.
Make sure the desired setting is displayed. is screen will show
whether the Upstream Occlusion Sensor is turned on or o.
WARNING: When the Upstream Occlusion Sensor is turned o,
the pump will not detect occlusions upstream (between pump and
uid container). Periodically inspect theuid container for de-
creasing volume, inspect theuid path for kinks, a closed clamp,
or other upstream occlusions. Upstream occlusions could result in
under- or non-delivery of medications. If undetected, these occlu-
sions could result in death or serious injury to the patient.
If you need to change the Upstream Occlusion Sensor setting, see
Section 4, Biomed Functions.
Press
.
12. Review the program.
Press
repeatedly to review the programming screens. If you
need to reprogram a setting, press
until the appropriate
screen appears and change the setting as described in this section.