EasyManua.ls Logo

CADD Legacy Plus 6500 - Annual Functional Inspection and Testing Procedures; Inspection Procedures

CADD Legacy Plus 6500
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
72
Reference
Section 5: Reference
Annual Functional Inspection and Testing Procedures
Smiths Medical MD, Inc. recommends annual functional inspections
and tests on all CADD-Legacy® pumps. e following inspection and
testing procedures should be performed annually to verify function and
accuracy. e pump must be in Lock Level 0 (LL0) to perform the follow-
ing inspections and tests.
NOTE: Persons performing the following tests and procedures should be
familiar with the CADD-Legacy® pump. Please read the entire Operators
Manual before proceeding.
CAUTION: CADD-Legacy® pumps are sealed units. A broken or
damaged seal will, therefore, be considered conclusive evidence that
the pump has been misused and/or altered, which voids any and all
warranties. All service and repair of CADD-Legacy® pumps must be
performed by Smiths Medical MD, Inc. or its authorized agents.
Inspection Procedures
Visual Inspection
Visually inspect the pump for any damage to the LCD, occlusion sen-
sor seals, valves and expulsor, pump hinge area, lock, cassette sensor,
keypad, indicator light, power jack, accessory jack, air detector, and
housing.
Check the battery door for proper operation. It should not be broken
or damaged. e mating tabs on the pump housing should not be
broken or damaged.
Examine the battery compartment for damage. If the battery contacts
appear corroded, clean them with a cotton swab and isopropyl alcohol
(see Cleaning the Battery Contacts). If the battery contacts appear
to be bent or pushed in, straightening may be possible with a small
screwdriver or other suitable tool. Care must be taken not to damage
the pump housing or to incur further damage to the contacts.

Table of Contents

Related product manuals