72
Reference
Section 5: Reference
Annual Functional Inspection and Testing Procedures
Smiths Medical MD, Inc. recommends annual functional inspections
and tests on all CADD-Legacy® pumps. e following inspection and
testing procedures should be performed annually to verify function and
accuracy. e pump must be in Lock Level 0 (LL0) to perform the follow-
ing inspections and tests.
NOTE: Persons performing the following tests and procedures should be
familiar with the CADD-Legacy® pump. Please read the entire Operator’s
Manual before proceeding.
CAUTION: CADD-Legacy® pumps are sealed units. A broken or
damaged seal will, therefore, be considered conclusive evidence that
the pump has been misused and/or altered, which voids any and all
warranties. All service and repair of CADD-Legacy® pumps must be
performed by Smiths Medical MD, Inc. or its authorized agents.
Inspection Procedures
Visual Inspection
• Visually inspect the pump for any damage to the LCD, occlusion sen-
sor seals, valves and expulsor, pump hinge area, lock, cassette sensor,
keypad, indicator light, power jack, accessory jack, air detector, and
housing.
• Check the battery door for proper operation. It should not be broken
or damaged. e mating tabs on the pump housing should not be
broken or damaged.
• Examine the battery compartment for damage. If the battery contacts
appear corroded, clean them with a cotton swab and isopropyl alcohol
(see Cleaning the Battery Contacts). If the battery contacts appear
to be bent or pushed in, straightening may be possible with a small
screwdriver or other suitable tool. Care must be taken not to damage
the pump housing or to incur further damage to the contacts.