Section 5: Reference
Messages and Alarms Description/Corrective Action
High Pressure
Two-Tone alarm
e pump has detected high pressure,
which may be resulting from a
downstream blockage, kink in the uid
path, or a closed tubing clamp. Remove
the obstruction to resume operation.
Or, press
or
to stop the pump
and silence the alarm for 2 minutes, then
remove the obstruction and restart the
pump.
Key pressed,
Please release
Two-Tone alarm
If a key is being pressed, stop pressing
it. If the alarm persists, close the tubing
clamp and remove the pump from use.
Contact Customer Service to return the
pump for service.
LowBat
Three Two-Tone beeps
every 5 minuTes
e batteries are low, but the pump is
still operable.
Change the batteries soon.
Motor Locked,
remove all power
Two-Tone alarm
Batteries are depleted and the pump was
powered up with the AC Adapter. Install
new AA batteries, reconnect the AC
adapter, and restart the pump.
[No message]
Two-Tone alarm
With no AC adapter attached, the batter-
ies have been removed while the pump is
running. e pump is now stopped and
unpowered. Install batteries to silence the
alarm.
OR
Batteries were removed within
approximately 15 seconds aer stopping the
pump. Install new batteries to silence the
alarm, if desired. Otherwise, the alarm will
stop within a short period of time.