Section 2: Pump Setup and Programming
Pump Setup &
Programming
22
Continuous Rate
Enter the continuous rate of medication delivery (in ml/hr). e maxi-
mum rate is 125 ml/hr.
Given
is screen shows the total amount of medication delivered since the last
time the value was cleared. e amount shown is rounded to the nearest
0.05 ml. If this value reaches 99999.95, it automatically returns to 0 and
continues counting. e Given amount does not include medication used
when priming the tubing.
Air Detector Status
is screen indicates whether the Air Detector is on high sensitivity, low
sensitivity or turned o. e Air Detector status cannot be changed with-
out entering the Biomed Functions Code (see Section 4, Biomed Functions,
to change the setting).
Upstream Sensor Status
is screen indicates whether the Upstream Occlusion Sensor is turned
on or turned o. e Upstream Sensor status cannot be changed without
entering the Biomed Functions Code (see Section 4, Biomed Functions, to
change the setting).
Before programming in Continuous Delivery Mode:
Stop the pump and change the Lock Level to LL0 as described in Section 1.
Select the Continuous Delivery Mode (CDM) as described in Chang-
ing the Delivery Methods in Section 4. If the pump is already in the
CDM, you may select it again to clear all programming screens to
their default settings.