Section 2: Pump Setup and Programming
Pump Setup &
Programming
24
WARNING: When the Air Detector is turned o, the pump will
not detect air in the uid path. Periodically inspect the uid
path and remove any air to prevent air embolism. Air embolism
could result in serious injury or death to the patient.
• If you need to change the Air Detector setting, see Section 4,
Biomed Functions.
• Press
„
.
6. Verify the Upstream Sensor status.
• Make sure the desired setting is displayed. is screen will show
whether the Upstream Occlusion Sensor is turned on or o.
WARNING: When the Upstream Occlusion Sensor is turned o,
the pump will not detect occlusions upstream (between pump
and uid container). Periodically inspect the uid container
for decreasing volume, inspect the uid path for kinks, a closed
clamp, or other upstream occlusions. Upstream occlusions could
result in under- or non-delivery of medications. If undetected,
these occlusions could result in death or serious injury to the
patient.
• If you need to change the Upstream Occlusion Sensor setting, see
Section 4, Biomed Functions.
• Press
„
.
7. Review the program.
Press
„
repeatedly to review the programming screens. If you
need to reprogram a setting, press
„
until the appropriate
screen appears and change the setting as described in this section.