Section 2: Pump Setup and Programming
Pump Setup &
Programming
34
WARNING: Attach the cassette (the part of the Medication Cas-
sette Reservoir or CADD™ Administration Set that attaches to
the pump) properly. An improperly attached or detached cas-
sette could result in unregulated gravity infusion of medication
from the uid container or a reux of blood, which could result
in death or serious injury to the patient.
If you are using a CADD™ Administration Set or Medication
Cassette Reservoir that does not have the ow stop feature
(catalog number does not start with 21-73xx): You must use a
CADD™ Extension Set with an integral Anti-Siphon Valve or a
CADD™ Administration Set with either an integral or Add On
Anti-Siphon Valve to protect against unregulated gravity infu-
sion that can result from an improperly attached cassette.
5. Gently twist, push, and pull on the cas-
sette to make sure it is rmly attached.
If the cassette is not secure, repeat the
procedure from step 1.