EasyManua.ls Logo

CAE Apollo - Page 78

CAE Apollo
97 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
©2020 CAE 905K351052 v1.1 73
Care and Maintenance
f. Reassemble the in-line filter, ensuring the black rubber seal is in place at the base of the
blue filter cone.
Troubleshooting the Trauma Fill Tank
Before making any repairs, ALWAYS depressurize the tank, remove all trauma solution and clean the
tank.
Empty and Flush the Chest Tube Reservoir
Removing fluids from the Chest Tube reservoir and the Chest Tube system requires the same steps.
Problem Cause Solution
Tank can be pressurized, but
only air comes out.
Siphon tube has detached
from insert.
Remove hose from tank and
reinsert siphon tube.
Pressure does not build up. No
fluid is transported to simula-
tor.
Pump assembly not sealed
tightly into tank
OR
Damaged pump cylinder
gasket or o-ring
OR
Tank pressure relief valve is set
to “open.”
Thoroughly clean pump cylin-
der gasket or o-ring and sur-
rounding area and apply a light
coating of silicone to pump
gasket or o-ring.
Contact CAE for service.
Turn valve until it returns to a
“sealed” position.
Simulator fill time is too long
(more than 5 minutes).
Not enough strokes applied to
create pressure
OR
The in-line filter is dirty
OR
The umbilical is disconnected
at Overflow Bottle or (4) Too
much fluid in fill tank.
Pump 25 to 35 times for best
performance.
Clean filter.
Reconnect the overflow fitting.
Note: The Trauma Fill Tank
works best with 1 gallon (3.6
liters) of fluid inside. If greater
amounts of fluid are used, tank
may require additional pumps
as fluid is transported to simu-
lator.

Table of Contents

Related product manuals