©2021 CAE 905K690052 v1.0 69
Using Aria
To defibrillate using SymDefib:
1. Connect the SymDefib cable to the defibrillator using the cables provided. The SymDefib
system supports Phillips defibrillators without a cable adapter. If using Zoll or Physio-
Control defibrillators, you must use the appropriate adapter between the cable on the
SymDefib and the defibrillator.
Connecting Power Adapter to the SymDefib
2. If using a power supply (instead of the battery), plug the SymDefib power cord into the
back port of the SymDefib box and plug the other end into the power source.
3. Connect the electrode pad cable to the right port on the front side of the SymDefib box.
4. With the manikin started, turn on the SymDefib and verify the status light flashes green to
indicate it is on.
5. While the SymDefib is operating, attach the electrode pads to the magnetic connectors on
the manikin. When connected, the SymDefib begins to display an ECG signal on the
monitor and the status light turns a solid green.
Note: The SymDefib was paired to the simulator in the factory and should automatically connect
wirelessly.