CAE Aria - Step 8: Power off the Instructor Tablet; Step 9: Power off the Manikin

CAE Aria
102 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
©2021 CAE 905K690052 v1.0 91
Care and Maintenance
4. Run a Bleeding SCE simulation in Maestro or turn on the Flush command located on the
maintenance tab in Maestro.
5. Continue to run the SCE until all clean distlled water is pushed through the circuit.
6. Repeat steps 3 through 5 for the other bleeding port.
To drain the bleeding system:
1. When the external bleed tank empties, air will be pushed through the circuit.
2. Allow the air to pass, removing all fluid, one port at a time.
3. Stop the SCE simulation in Maestro or turn off the Flush command located on the
maintenance tab in Maestro.
Step 8: Power off the Instructor Tablet
To power off the Surface Go 2 instructor tablet:
1. Tap the white back arrow at the top left of the Maestro run screen to exit the current
simulation session.
2. Tap Stop and Exit.
3. Tap the X in the upper-right corner of the screen to exit Maestro.
4. The tablet desktop screen appears.
5. Tap the Start menu icon in the lower-left corner of the desktop screen.
6. Tap Power and then tap Shut Down.
Step 9: Power Off the Manikin
To power off the manikin:
1. Press and release the Power button located on the bottom of the manikin’s right foot.
The LED will flash Green for approximately 45 seconds, then turn off when the unit has completely
powered down. After the unit is powered down, the LED will blink Blue approximately every 30
seconds indicating the unit is in low power mode.
Note: The power button will flash blue every 30 seconds to indicate the low power mode is enabled.
If the button no longer flashes, the battery has been turned off and external power will need to be