Setup
30 ©2021 CAE 905K690052 v1.0
CAEAria
Step 10: Prime the Fluids Systems
To prime the Antecubital IV site:
1. Attach the IV tubing and empty IV bag to the drain port on the manikin. Make sure the
valve on the spike set is open.
2. Remove cap from the antecubital fill port on the back of the left upper arm.
3. Fill sryinge with 20 ml of distillled fluid and attach empty syringe to the IV fill port.
Note: You may use distilled fluid mixed with red food coloring to simulate flashback if
desired.
4. Inject the 20 ml of distilled water and remove the syringe and replace the fill port cap.
Priming the IV System for Flashback
5. The antecubital site is now ready for venipuncture.
To prime the jugular and left hand preported IV sites:
1. Attach the IV tubing and empty IV bag to drain port on the manikin.
2. Remove cap on the IV port sit and inject 10 mL of fluid. Replace cap.
3. During simulation, attach the IV tubing and fluids to be infused.
4. Begin fluid infusion.
IV Port