Care and Maintenance
86 ©2021 CAE 905K690052 v1.0
CAEAria
Breakdown
To ensure your simulator remains in good working condition, follow the CAE breakdown procedures
below prior to storing.
Step 1: Clean the Manikin
To clean the simulator, use a soft cloth and warm water with a mild detergent to wipe off any
moulage materials on the manikin’s skin. Gently rub the soiled area with a soft cloth to remove most
marks and stains. DO NOT use solvents or abrasive pads.
Inspect the manikin for damage or puncture marks. If damage exists, contact CAE Healthcare
Customer Service immediately for a repair. Prompt repair prevents expansion of the damaged area.
Step 2: Flush and Drain the Left Arm IV Sites
To flush the Antecubital and Dorsum of the left Hand IV sites:
1. Attach the IV tubing and empty IV bag to the fluid drain port located at the manikin’s
buttocks.
2. If non-clear distilled water was used to prime the site or infused:
a. Remove the antecubital IV prime port cap and attach syringe filled
with 50 ml of clear distilled water.
b. Flush with clear distilled water until clear water exits from the drain
port.
c. Remove syringe and replace IV port cap.
d. Repeat steps a through c for the left hand site.
Break down the Aria Simulator
1
Clean the Manikin.
2
Flush and Drain the Left Arm IV sites.
3
Maintain IM Injection Site.
4
Flush and Drain the Jugular, Tibial IO and Gastrostomy Sites.
5
Flush and Drain the Urinary System.
6
Flush and Drain the Chest Tube.
7
Flush and Drain the Bleeding System.
8
Power Off the Instructor Tablet.
9
Power Off the Manikin.