ADC 2
Instruction Manual, November 2016 Page 12
4. Manual operation
4.1. HPTLC plate / plate holder / foil holder
The plate-holder is designed to attach and handle the HPTLC
plates in the ADC2. When inserting a plate into the plate holder
the layer side must be facing the user (‘LAYERside).
Fig. 5: Insert TLC plate in plate-holder
Important: While clamping tight the TLC plate the user
should be aware the plate fits the inside stop position and
the edges of holder and plate do not overlap.
Fig. 6: Plate clamped in Plate-holder
Important: When clamping the HPTLC plate make sure the
plate has hit the stop position (1) and the edges of the
holder (2) do not overlap the glass plate.
1
2