2.2. Installation environment
The place for installation must meet the following requirements:
Bench space Width 320mm
Depth 320mm without Humidity Option
Depth 460mm with Humidity Option
(additional for tubes 250mm)
Height 520mm
Weight 18.5kg
Operating
temperature
The temperature should be within a range of 18 to 30
degrees centigrade and free from significant variations.
Humidity
Humidity and temperature conditions must not cause
condensation.
Atmospheric
conditions
Adequate ventilation free from acidic, alkaline or other
gas that may corrode metal or painted
surfaces must be
secured.