Further requirements:
Do not place the instrument in a location where the tempera-
ture significantly changes (e.g. under an air conditioning duct
or by a window). Significant changes in temperature will affect
the performance of the unit
Do not place the instrument in direct sunlight. Direct sunlight
will disturb the measurement of the CCD sensor
Do not use the instrument in an environment with moving am-
bient air (draft)
Do not expose the instrument to any strong vibration or shock
Avoid placing the instrument near equipment that radiates
heat. Do not place the instrument near gas burners, electric
heaters or ovens
Do not place the instrument near equipment that generates in-
tense magnetic fields such as electric welding equipment, high
frequency furnaces, pole transformers, etc.
Protect the instrument from excessive dust
Connect the instrument to power lines that are free from sud-
den changes or voltage fluctuations
If you must use power motor driven equipment (such as a stir-
rer or shaker) in the same line as your instrument, ensure that a
noise reduction unit is in the same power line
2.3. Conditions for the installation
Confirm that the following requirements exist before installing the
instrument:
Power supply and ground
Line voltage: 100 240V (see rating plate on instrument).
Frequency: 50 / 60 Hz
Power capacity: 20 W
Ground terminal:
A grounded outlet should be located within 2 me-
ters of the instrument.