CALIBRATION and SETUP, Cont.
OPERATIONAL SETUP, Cont.
PRINT SETUP
The display will show prsetprset, the prompt to enable printing a record of the scale's current
Operational Setup and display the current (default) setting NO. This operation, when enabled
and with a printer connected to the printer port will print a summary of the Operational Setup.
To print the summary, press the YES/1 key, then press he ENTER key. The summary will
begin to print immediately after pressing the ENTER key. Upon completion of printing, the
setup prompt will automatically return to NO, the default setting and then proceed to the next
prompt.
Please note, that in order to print the Operational Setup, a serial dot matrix printer and
8½" x 11" paper must be used NOT the thermal label printer used for printing labels.
The serial dot matrix printer must have a RS232 serial interface configured for 9600
baud, 8 data bits, No parity and 1 stop bit (9600,8,N,1).
SETUP DONE
If Setup was performed using the Calibration Method (sealing screw REMOVED), the display
will change to setup doNE setup doNE when completed. Place the power switch in the OFF position
and replace the calibration sealing screw on the right side panel of the scale. Return the
power switch to the ON position. The scale will power up to the Weight Display mode and is
now ready for use.
When using the Operation Method (sealing screw INSTALLED), at the completion of Setup,
the scale will return to Weight Display mode and is ready for use.
CALIBRATION PROCEDURE - REMOTE SCALE
If sealing wires require breaking for purposes of calibration, proper procedures under
National Institue of Standards and Technology Handbook 44 must be adhered to.
1. With the scale power off, remove the calibration sealing screw on the right side panel, see
Figure No.7.
2. With the screw removed, place the power switch in the ON position. The scale will perform a
brief lamp test. This test consists of illuminating all display segments and annunciator LED's
for approximately three (3) seconds to allow the operator to make a visual verification that
the display is operational. After completion of the lamp test, the scale will display the model
number and software revision level, then the weight display will change to SCALESCALE.
3. Select the REMOTE scale by pressing the REMOTE SCALE key.
NOTE: The REMOTE scale cannot be selected until it has first been enabled in
Setup (refer to Operational Setup section).
4. The weight display will change to cap =cap = (capacity).
5. Key-in the scale capacity (max 6 digits) and press the ENTER key.
6. The display will change to int =int = (interval or division).
7. Key-in the interval (division value) (max 4 digits) then press the ENTER key. The following
values are allowable for the interval: 1, 2, 5, .1, .2, .5, .01, .02, .001, .002, .005, .0001,
.0002 and .0005.
29