INTERCONNECTIONS
Remote Scale Connection
If a remote scale is used, it connects to the counting scale via a 9-pin "D" connector (REMOTE
SCALE) located on the rear panel of the scale (see Figure No. 2). Before connecting your
remote scale to the counting scale, verify that your scale cable is correctly wired. Refer to Figure
No. 3 for pin identification of the remote scale connector.
The Digital Counting Scale ships from the factory configured for remote scales that do not use
load cells with sense leads (4 wire load cells). The sense jumpers, J1 and J2 are installed for
proper operation with 4 wire load cells. If your remote scale uses load cells with sense leads (6
wire load cells), jumpers J1 and J2, located on the printed circuit board (see Figure No. 5) should
be placed on one pin only or disconnected.
Make certain that the connector retaining screws are used to hold the remote scale cable
connector securely to the rear panel.
Figure No. 3
MATING CONNECTOR INFORMATION
PART NO. VENDOR PART NO.
6610-2379 DE-9P CONNECTOR
6610-1131 DE-24657 SHELL
6
REMOTE SCALE
PIN LOCATION AS VIEWED FROM CABLE ATTACHMENT SIDE
1
69
5