PRECAUTIONS
Most scales are designed for an office type environment. The Digital Counting Scale is no
exception to that rule. Such an environment is free of excessive dust and moisture and provides
a comfortable temperature. In general, the scale will perform well over a temperature range of
32° to 104° F (0° to +40° C). In order to keep cooling requirements to a minimum, the scale
should be placed out of direct sunlight and to provide adequate air circulation, keep the area
around the scale clear. Make certain the scale is not directly in front of a heating or cooling vent.
Such a location will subject the scale to sudden temperature changes and air currents which may
result in unstable weight readings. Insure that the scale has good, clean AC power and is
properly grounded. In areas subject to lightning strikes, additional protection to minimize
lightning damage, such as surge suppressors, should be installed.
5
CAUTION! When in parallel runs, locate external Load Cell
cables a minimum of 24 inches away from all AC wiring.