CS 1200 Intraoral Camera User and Installation Guide (9H1809)_Ed04 31
8 Maintenance
Perform the following maintenance activities on your camera and accessories regularly.
The camera is not delivered in disinfected condition. To ensure maximum hygienic safety for the patient,
carefully follow the instructions to disinfect the camera before use (see “Cleaning and Disinfecting the
Camera” ).
Cleaning and Disinfecting the Camera
General Warnings
WARNINGS:
Read and follow the warnings and personal protection
instructions provide in the Material Safety Data Sheet
(MSDS) of the disinfectant before use.
The disinfectant should be approved by the applicable
competent authority for use on a dental medical device.
You must wear gloves while handling and using the camera.
Always protect the camera with a hygienic barrier sheath
before use.
The camera must be disinfected with an EPA-registered
intermediate-level disinfectant solution with tuberculocidal
activity between patients.
DO NOT use a disinfectant containing phenolics or
iodophors, doing so will damage the surface coating of the
camera.
Never put the camera in a sterilizing device or immerse it
completely.
Excessive fluids can damage the camera.
Not protected against water spray.