Specification
2-1
2. Specification (15°C~ 35°C RH ≤ 75%)
2.1 Measurement Function
Primary Parameter:
L Inductance Unit: uH, mH, H, KH
C Capacitance Unit: pF, nF, uF, mF
R Resistance Unit: m, , K, M
|Z| Absolute value of impedance Unit: m, , K, M
Secondary Parameter:
Q Quality factor
D Dissipation factor
ESR Equivalent series resistance Unit: m, , K, M
Phase angle Unit: ∘ (degree)
Xs Reactance Unit: m, , K, M
Equivalent Measurement Circuit: Parallel, Serial
Range: Auto, Manual
Trigger Mode: Internal, Manual and External (GPIB, RS-232,
Handler Interface)
Measurement Terminal: 4 terminals test
Measurement Speed: Fast, Medium and Slow
2.2 Test Signal
Frequency:
11021 : 100Hz, 120Hz, 1kHz, 10kHz (the actual output frequency is 9.6KHz)
Accuracy : ± 0.25%
11021-L : lKHz, 10KHz, 40KHz, 50KHz
Accuracy : ± 0.02%
Voltage: Constant voltage: 11021: 0.25Vrms, 1V rms
11021-L: 50mVrms, 1V rms
Output Impedance: 25, 100, 1K, 10K, 100K depends on the impedance.